Physik - Open Access LMU - Teil 02/02
Physik - Open Access LMU - Teil 02/02
Ludwig-Maximilians-Universität München
Accuracy of linear depolarisation ratios in clear air ranges measured with POLIS-6 at 355 and 532 nm 1/2
1 seconds Posted Jan 1, 2015 at 11:00 am.
0:00
0:01
Download MP3
Show notes
Linear depolarization ratios in clean air ranges were measured with POLIS-6 at 355 and 532 nm. The mean deviation from the theoretical values, including the rotational Raman lines within the filter bandwidths, amounts to 0.0005 at 355 nm and to 0.0012 at 532 nm. The mean uncertainty of the measured linear depolarization ratio of clean air is about 0.0005 at 355 nm and about 0.0006 at 532 nm.